Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Wavetek DM23XT: (1 year) CAL VER /5520 INSTRUMENT: Wavetek DM25XT: (1 year) CAL VER /5520 INSTRUMENT: Wavetek DM27XT: (1 year) CAL VER /5520 DATE: 12-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 75 NUMBER OF LINES: 246 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP Procedure not formally tested with UUT. 1.001 DISP See procedure README file. 1.002 ASK- P F W 1.003 ASK+ K X 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Warm-up time: 30 min. - REQUIRED 1.005 DISP Temperature: 20 - 30 degC 1.005 DISP Humidity: <80% 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 HEAD {DC VOLTAGE} 1.008 DISP Connect the 5520A to the UUT as follows: 1.008 DISP [32] 5520A NORMAL HI to UUT VOLT/OHM 1.008 DISP [32] 5520A NORMAL LO to UUT COM 1.009 HEAD DC VOLTAGE: {200mV Range} 1.010 5520 200 190.0mV 0.5% 0.1U 2W 2.001 HEAD DC VOLTAGE: {2V Range} 2.002 5520 2 1.900V 0.5% 0.001U 2W 3.001 HEAD DC VOLTAGE: {20V Range} 3.002 5520 20 19.00V 0.5% 0.01U 2W 4.001 HEAD DC VOLTAGE: {200V Range} 4.002 5520 200 190.0V 0.5% 0.1U 2W 5.001 HEAD DC VOLTAGE: {1000V Range} 5.002 5520 1000 900V 0.5% 1U 2W 6.001 HEAD {AC VOLTAGE} 6.002 HEAD AC VOLTAGE: {200mV Range} 6.003 5520 200 190.0mV 1% 0.4U 50H SI 2W 7.001 5520 200 190.0mV 1% 0.4U 400H SI 2W 8.001 5520 200 190.0mV 1.5% 0.4U 1kH SI 2W 9.001 HEAD AC VOLTAGE: {2V Range} 9.002 5520 2 1.900V 1% 0.004U 50H SI 2W 10.001 5520 2 1.900V 1% 0.004U 400H SI 2W 11.001 5520 2 1.900V 1.5% 0.004U 1kH SI 2W 12.001 HEAD AC VOLTAGE: {20V Range} 12.002 5520 20 19.00V 1% 0.04U 50H SI 2W 13.001 5520 20 19.00V 1% 0.04U 400H SI 2W 14.001 5520 20 19.00V 1.5% 0.04U 1kH SI 2W 15.001 HEAD AC VOLTAGE: {200V Range} 15.002 5520 200 190.0V 1% 0.4U 50H SI 2W 16.001 5520 200 190.0V 1% 0.4U 400H SI 2W 17.001 5520 200 190.0V 1.5% 0.4U 1kH SI 2W 18.001 HEAD AC VOLTAGE: {750V Range} 18.002 5520 750 700V 1% 4U 50H SI 2W 19.001 5520 750 700V 1% 4U 100H SI 2W 20.001 5520 750 700V 1.5% 4U 1kH SI 2W 21.001 HEAD {RESISTANCE} 21.002 HEAD RESISTANCE: {2000 MOhm Range} 21.003 5520 2000 300.0MZ 10% 10U 2W 22.001 HEAD RESISTANCE: {20 MOhm Range} 22.002 5520 20 19.00MZ 2% 0.05U 2W 23.001 HEAD RESISTANCE: {2 MOhm Range} 23.002 5520 2 1.900MZ 1% 0.004U 2W 24.001 HEAD RESISTANCE: {200 kOhm Range} 24.002 5520 200 190.0kZ 1% 0.4U 2W 25.001 DISP Connect the 5520A to the UUT as follows: 25.001 DISP [32] 5520A AUX HI to UUT V/OHM 25.001 DISP [32] 5520A AUX LO to UUT COM 25.001 DISP [32] 5520A NORMAL HI to UUT V/OHM 25.001 DISP [32] 5520A NORMAL LO to UUT COM 25.001 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 25.001 DISP the NORMAL leads to the SENSE leads AT the UUT. 25.002 HEAD RESISTANCE: {200 Ohm Range} 25.003 5520 200 190.0Z 1% 0.4U CW 26.001 HEAD RESISTANCE: {2 kOhm Range} 26.002 5520 2 1.900kZ 1% 0.004U CW 27.001 HEAD RESISTANCE: {20 kOhm Range} 27.002 5520 20 19.00kZ 1% 0.04U CW 28.001 HEAD {CONTINUITY} 28.002 DISP Set the UUT to Diode/Continuity mode. 28.003 HEAD CONTINUITY: {400 Ohms} 28.004 5520 400.0Z S CW 28.005 EVAL Is {continuity alarm off}? 29.001 HEAD CONTINUITY: {<75 Ohms} 29.002 5520 70Z S CW 29.003 EVAL Is {continuity alarm sounding}? 30.001 HEAD {DIODE} 30.002 DISP Connect the 5520A to the UUT as follows: 30.002 DISP [32] 5520A NORMAL HI to UUT V/OHM 30.002 DISP [32] 5520A NORMAL LO to UUT COM 30.003 5520 0V S 2W 30.004 EVAL [N]Does {display indicate an overload condition}? 31.001 5520 0.700V 1.5% 0.1U 2W 32.001 JMPT 52.001 Wavetek DM23XT: (1 year) CAL VER /5520 32.002 HEAD {CAPACITANCE} 32.003 JMPT 39.002 Wavetek DM27XT: (1 year) CAL VER /5520 32.004 MATH MEM=M[1] 32.005 DISP Connect 5520A to the UUT as follows: 32.005 DISP [32] 5520A NORMAL HI to UUT mA Cx 32.005 DISP [32] 5520A NORMAL LO to UUT COM 32.006 HEAD CAPACITANCE: {2nF Range} 32.007 5520 2 1.900nF 2% 0.004U 2W 33.001 HEAD CAPACITANCE: {20nF Range} 33.002 5520 20 19.00nF 2% 0.04U 2W 34.001 DISP Connect the 5520A to the UUT as follows: 34.001 DISP [32] 5520A AUX HI to UUT mA Cx 34.001 DISP [32] 5520A AUX LO to UUT COM 34.001 DISP [32] 5520A NORMAL HI to UUT mA Cx 34.001 DISP [32] 5520A NORMAL LO to UUT COM 34.001 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 34.001 DISP the NORMAL leads to the SENSE leads AT the UUT. 34.002 HEAD CAPACITANCE: {200nF Range} 34.003 5520 200 190.0nF 2% 0.4U CW 35.001 HEAD CAPACITANCE: {2uF Range} 35.002 5520 2 1.900uF 2% 0.004U CW 36.001 HEAD CAPACITANCE: {20uF Range} 36.002 5520 20 19.00uF 2% 0.04U CW 37.001 HEAD CAPACITANCE: {200uF Range} 37.002 5520 200 190.0uF 2% 0.4U CW 38.001 HEAD CAPACITANCE: {2000uF Range} 38.002 5520 2000 1000uF 2% 4U CW 39.001 JMP 44.001 39.002 DISP Connect 5520A to the UUT as follows: 39.002 DISP [32] 550A NORMAL HI to UUT mA Cx Lx 39.002 DISP [32] 550A NORMAL LO to UUT COM 39.003 HEAD CAPACITANCE: {2nF Range} 39.004 5520 2 1.900nF 5% 0.01U 2W 40.001 HEAD CAPACITANCE: {20nF Range} 40.002 5520 20 19.00nF 5% 0.1U 2W 41.001 DISP Connect the 5520A to the UUT as follows: 41.001 DISP [32] 5520A AUX HI to UUT mA Cx Lx 41.001 DISP [32] 5520A AUX LO to UUT COM 41.001 DISP [32] 5520A NORMAL HI to UUT mA Cx Lx 41.001 DISP [32] 5520A NORMAL LO to UUT COM 41.001 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 41.001 DISP the NORMAL leads to the SENSE leads AT the UUT. 41.002 HEAD CAPACITANCE: {2uF Range} 41.003 5520 2 1.900uF 5% 0.01U CW 42.001 HEAD CAPACITANCE: {20uF Range} 42.002 5520 20 19.00uF 5% 0.1U CW 43.001 HEAD CAPACITANCE: {2000uF Range} 43.002 5520 2000 1000uF 5% 10U CW 44.001 HEAD {FREQUENCY} 44.002 DISP Connect the 5520A to the UUT as follows: 44.002 DISP [32] 5520A NORMAL HI to UUT VOLT/OHM 44.002 DISP [32] 5520A NORMAL LO to UUT COM 44.003 DISP Set the UUT to Frequency mode. 44.004 JMPT 47.002 Wavetek DM27XT: (1 year) CAL VER /5520 44.005 5520 60H 0.5% 3U 1V SI 2W 45.001 5520 100H 0.5% 3U 1V SI 2W 46.001 5520 1kH 0.5% 3U 1V SI 2W 47.001 JMP 52.001 47.002 5520 60H 0.5% 3U 1V SI 2W 48.001 5520 100H 0.5% 3U 1V SI 2W 49.001 5520 1kH 0.5% 3U 1V SI 2W 50.001 5520 10kH 0.5% 3U 1V SI 2W 51.001 5520 1MH 0.5% 3U 1V SI 2W # 20MHz range not checked on DM27XT. 52.001 HEAD {DC CURRENT} 52.002 RNG 200uA 52.003 DISP Connect 5520A to the UUT as follows: 52.003 DISP [32] 5520A AUX HI to UUT mA 52.003 DISP [32] 5520A AUX LO to UUT COM 52.004 HEAD DC CURRENT: {200uA Range} 52.005 5520 200 190.0uA 1% 0.1U 2W 53.001 HEAD DC CURRENT: {2mA Range} 53.002 5520 2 1.900mA 1% 0.001U 2W 54.001 HEAD DC CURRENT: {20mA Range} 54.002 5520 20 19.00mA 1% 0.01U 2W 55.001 HEAD DC CURRENT: {200mA Range} 55.002 5520 200 190.0mA 1% 0.1U 2W 56.001 HEAD {AC CURRENT} 56.002 HEAD AC CURRENT: {200uA Range} 56.003 5520 200 190.0uA 1.5% 0.4U 100H SI 2W 57.001 HEAD AC CURRENT: {2mA Range} 57.002 5520 2 1.900mA 1.5% 0.004U 100H SI 2W 58.001 HEAD AC CURRENT: {20mA Range} 58.002 5520 20 19.00mA 1.5% 0.04U 100H SI 2W 59.001 HEAD AC CURRENT: {200mA Range} 59.002 5520 200 190.0mA 1.5% 0.4U 100H SI 2W 60.001 JMPT 62.002 Wavetek DM25XT: (1 year) CAL VER /5520 60.002 JMPT 62.002 Wavetek DM27XT: (1 year) CAL VER /5520 60.003 HEAD AC CURRENT: {10A Range} 60.004 RNG 10A AC 60.005 DISP Connect the 5520A to the UUT as follows: 60.005 DISP [32] 5520A 20A to UUT 10A 60.005 DISP [32] 5520A AUX LO to UUT COM 60.006 5520 10 10.00A 2.5% 0.04U 100H SI HC 2W 61.001 HEAD {DC CURRENT} 61.002 HEAD DC CURRENT: {10 A Range} 61.003 5520 10 10.00A 2% 0.03U HC 2W 62.001 JMP 64.002 62.002 HEAD AC CURRENT: {20 A Range} 62.003 RNG 20A AC 62.004 DISP Connect 5520A to the UUT as follows: 62.004 DISP [32] 5520A 20A to UUT 20A 62.004 DISP [32] 5520A AUX LO to UUT COM 62.005 5520 20 10.00A 2.5% 0.04U 100H SI HC 2W 63.001 HEAD {DC CURRENT} 63.002 HEAD DC CURRENT: {20 A Range} 63.003 5520 20 10.00A 2% 0.03U HC 2W 64.001 JMP 75.001 64.002 HEAD {TEMPERATURE} 64.003 DISP Connect the 5520A TC terminals to the UUT TC + and - 64.003 DISP Inputs using type K thermocouple wire and connectors. 64.004 HEAD TEMPERATURE: {200 degC Range} 64.005 5520 200 -19.0degC 1% 2U _K TC 2W 65.001 5520 200 0.0degC 1% 2U _K TC 2W 66.001 5520 200 190.0degC 1% 2U _K TC 2W 67.001 HEAD TEMPERATURE: {750 degC Range} 67.002 5520 750 200degC 3% 2U _K TC 2W 68.001 5520 750 700degC 3% 2U _K TC 2W 69.001 HEAD TEMPERATURE: {200 degF Range} 69.002 5520 200 -4.0degF 1% 0.4U _K TC 2W 70.001 5520 200 32.0degF 1% 0.4U _K TC 2W 71.001 5520 200 190.0degF 1% 0.4U _K TC 2W 72.001 HEAD TEMPERATURE: {1400 degF Range} 72.002 5520 1400 200degF 3% 4U _K TC 2W 73.001 5520 1400 1000degF 3% 4U _K TC 2W 74.001 5520 1400 1350degF 3% 4U _K TC 2W 75.001 END